Z-3225-2017Class II

FUJINON ED-530XT. This device is intended for the visualization of the duodenum and upper digestive tract...

FUJIFILM Healthcare Americas Corporation / initiated 2017-07-21 / Terminated / root cause: design

Fujifilm Medical Systems U.S.A., Inc. began a recall of FUJINON ED-530XT. This device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and... on . FDA classified it as Class II and gives the reason as "An update to the design and labeling was implemented to help reduce patient risk associated with inadequate reprocessing of the device. The action includes replacement of the forceps elevator mechanism, the O-ring seal...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-3225-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
FUJINON ED-530XT. This device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.
Product code
FDT Duodenoscope And Accessories, Flexible/Rigid
Reason for recall
An update to the design and labeling was implemented to help reduce patient risk associated with inadequate reprocessing of the device. The action includes replacement of the forceps elevator mechanism, the O-ring seal, the distal end cap and issuance of a new Operation Manual.
Root cause tag
design
FDA root cause
Device Design
Action
On July 21, 2017, Fujifilm Medical Systems U.S.A., Inc. Endoscopy Division distributed Urgent Medical Device Correction and Removal notices & Field Action Verification Forms to their U.S. Consignees by USPS certified mail. Fujifilm Medical Systems will contact all customers that have the duodenoscope by October 2017, with details for returning the device for replacement parts. This field action allows continued use of the current device until you are contacted by a sales representative for replacement parts as early as October 2017. While the remediation of the scopes will take approximately two weeks to complete, Fujifilm Medical Systems U.S.A. Inc. Endoscopy Division will provide interim duodenoscopes on loan for continuity of care. Once received, the recalling firm will replace the forceps elevator mechanism with O-ring and the distal end cap. An updated manual has been generated to include new information/guidance. Customers are encouraged to complete and return the Field Action Verification Form via email to [email protected] or via fax to (973) 872-4723. Customers with questions can call Larry Picciano, Sr. Director Quality and Regulatory Affairs at (973) 686-2479.
Quantity in commerce
362 units
Distribution
Nationwide
Product codes and lots
All lots/serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3225-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3225-2017Class IITerminatedVoluntary: Firm initiated