PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
Handicare Usa Inc began a recall of PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered on . FDA classified it as Class II and gives the reason as "During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0013-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Handicare USA, Inc.
- Product
- PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
- Product code
- FSA Lift, Patient, Non-Ac-Powered
- Reason for recall
- During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- The firm initiated their recall by email on 09/08/2017. The consignee was warned that the lift straps may wear prematurely and they would be contacted by a representative and the units would be replaced.
- Quantity in commerce
- 8 units
- Distribution
- Us and Canada
- Product codes and lots
- A6M0001526, A6M0001527, A6M0001528, A6M0001529, A6M0001530, A6M0001485, A6M0001486, A6M0001498
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0013-2018 | Class II | Terminated | Voluntary: Firm initiated |