Z-0033-2018Class II

Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and...

Abbott Medical / initiated 2017-08-28 / Terminated / root cause: software

St Jude Medical Inc. began a recall of Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component... on . FDA classified it as Class II and gives the reason as "New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0033-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broadband Kit. The transmitted data enables clinicians to remotely monitor a patients IMD using a secure, encrypted, password-protected web site, thus reducing the number of in-clinic visits a patient needs to make to their physician. The transmitter monitors the patients device daily between scheduled follow-ups and can alert a clinician if it detects anything out of the norm.
Product code
OSR Pacemaker/Icd/Crt Non-Implanted Components
Reason for recall
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
Root cause tag
software
FDA root cause
Unknown/Undetermined by firm
Action
Abbott sent an Important Cybersecurity Advisory dated August 28, 2017, to all affected customers to notify customers of the availability of the programmer software update and associated pacemaker firmware update. The notification material is in the form of a Physician Letter and Hospital Letter and will be delivered to physicians and hospitals with inventory in the US by overnight service. Customers with questions were instructed to contact their Abbott representative or the customers technical support hotline at 1-800-722-3774.
Quantity in commerce
495,011 units (391,093 in US)
Distribution
Worldwide Distribution
Product codes and lots
Model EX1100 - GTIN 05414734504782 and Model EX1150 - GTIN 05414734404334

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0033-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0029-2018Class IITerminatedVoluntary: Firm initiated
Z-0030-2018Class IITerminatedVoluntary: Firm initiated
Z-0031-2018Class IITerminatedVoluntary: Firm initiated
Z-0032-2018Class IITerminatedVoluntary: Firm initiated
Z-0033-2018Class IITerminatedVoluntary: Firm initiated
Z-0034-2018Class IITerminatedVoluntary: Firm initiated
Z-0035-2018Class IITerminatedVoluntary: Firm initiated
Z-0036-2018Class IITerminatedVoluntary: Firm initiated
Z-0037-2018Class IITerminatedVoluntary: Firm initiated
Z-0038-2018Class IITerminatedVoluntary: Firm initiated