THERMOCOOL SMARTTOUCH Bi-Directional Navigation Catheter (D-132704-S, D-132705-S)
Biosense Webster, Inc. began a recall of THERMOCOOL SMARTTOUCH Bi-Directional Navigation Catheter (D-132704-S, D-132705-S) on . FDA classified it as Class II and gives the reason as "Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0056-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biosense Webster Inc
- Product
- THERMOCOOL SMARTTOUCH Bi-Directional Navigation Catheter (D-132704-S, D-132705-S)
- Product code
- LPB Cardiac Ablation Percutaneous Catheter
- Reason for recall
- Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
- FDA root cause
- Under Investigation by firm
- Action
- All US and OUS customers, who have purchased the lots identified as part of this recall will be notified via customer letters in order to return the product to BWI facility where it will be segregated for further disposition. This product recall applies to hospitals (or direct customers) and distributors who have purchased and/or received the THERMOCOOL Catheters identified as part of this recall . One notification package which includes the Customer Notification Letter, Business Reply Form (B RF) and pre-paid return label will be distributed to the customers subject to this recall via a 3rd party vendor (Stericycle) for USA. OUS customer communication will occur in accordance with local regulations.
- Quantity in commerce
- 64 units
- Distribution
- US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China
- Product codes and lots
- 17445775M 17694994M 17697951M 17685286M 17697907M 17697991M 17691864M 17697910M 17697994M
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0054-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0055-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0056-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0057-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0058-2018 | Class II | Terminated | Voluntary: Firm initiated |