BARD MAX-CORE Disposable Core Biopsy Instrument, Product Codes - MC1410, 14 g x 10 cm; MC1616, 16 g x 16 cm...
Bard Peripheral Vascular Inc began a recall of BARD MAX-CORE Disposable Core Biopsy Instrument, Product Codes - MC1410, 14 g x 10 cm; MC1616, 16 g x 16 cm; MC1816, 18 g x 16 cm; MC1820, 18 g x 20 cm... on . FDA classified it as Class II and gives the reason as "During the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0044-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- BARD MAX-CORE Disposable Core Biopsy Instrument, Product Codes - MC1410, 14 g x 10 cm; MC1616, 16 g x 16 cm; MC1816, 18 g x 16 cm; MC1820, 18 g x 20 cm; MC1825, 18 g x 25 cm. The core needle biopsy device is intended for use in obtaining biopsies from soft tissues.
- Product code
- KNW Instrument, Biopsy
- Reason for recall
- During the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and Self-Activation.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- BPV sent a customer recall letter on September 22, 2017 to each of the US consignees via FedEx with proof of delivery notification. BPV will also initiate a recall for product outside the US in accordance with regulations of each country in which the product was distributed.
- Quantity in commerce
- 110,550 units
- Distribution
- US, Australia, Canada, Europe, Middle East, Africa, Latin America, Asia Pacific
- Product codes and lots
- Lot Numbers: REBP0019, REBN0342, REBN2123, REBP1199, REBP1419, REBP1807, REBQ0084, REBQ0343, REBQ1012, REBP1420, REBP1809, REBP0869, REBP1266, REBP1267, REBP1421, REBP1422, REBP1810, REBQ0087, REBQ0088, REBQ0347, REBQ0811, REBQ1014, REBP0158, REBQ1904, REBQ1898, REBQ1978, REBR0468, REBQ2296, REBR0474
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0044-2018 | Class II | Terminated | Voluntary: Firm initiated |