Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia...

Siemens Medical Solutions USA, Inc. / initiated 2017-09-18 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc. began a recall of Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model... on . FDA classified it as Class II and gives the reason as "The torque wrench used to tighten system bolts during installation was found to be out of tolerance". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0047-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
The torque wrench used to tighten system bolts during installation was found to be out of tolerance
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
The recall was initiated via phone and visit beginning 9/18/2017.
Quantity in commerce
16 devices
Distribution
Distribution was made to medical facilities located in CA, FL, GA, LA, NY, PA, and TN. There was government distribution but no foreign/military distribution.
Product codes and lots
Symbia T2 serial number 2001; Symbia Intevo Excel serial numbers 2112 and 2140; Symbia Intevo 16 serial numbers 2100 and 2116; Symbia Evo serial numbers 1049, 1054, 1071, 1075, 1104, and 1128; and Symbia Evo Excel serial numbers 1109, 1111, 1131, 1134, and 1158.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0047-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0047-2018Class IITerminatedVoluntary: Firm initiated