Z-0091-2018Class II

AQUIOS CL Flow Cytometer, Catalog No. B30166, Rx. The firm name on the label is Beckman Coulter Ireland Inc...

Siemens Healthcare Diagnostics Inc. / initiated 2017-10-02 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of AQUIOS CL Flow Cytometer, Catalog No. B30166, Rx. The firm name on the label is Beckman Coulter Ireland Inc., Ireland. The AQUIOS CL Flow Cytometer is intended... on . FDA classified it as Class II and gives the reason as "The system may duplicate sample requests leading to sample misidentification when connected to a Laboratory Information System (LIS)". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0091-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AQUIOS CL Flow Cytometer, Catalog No. B30166, Rx. The firm name on the label is Beckman Coulter Ireland Inc., Ireland. The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping. It is used in conjunction with the following reagents and software package: AQUIOS Tetra-1 Panel monoclonal antibody reagents, AQUIOS Tetra-2+ Panel monoclonal antibody reagents, and AQUIOS CL Flow Cytometry Software.
Product code
OYE Flow Cytometric Reagents And Accessories.
Reason for recall
The system may duplicate sample requests leading to sample misidentification when connected to a Laboratory Information System (LIS)
Root cause tag
software
FDA root cause
Software design
Action
Beckman Coulter sent an Urgent Medical Device Recall letter dated September 28, 2017, and via phone on October 4, 2017. On 11/30/2017, the firm issued an updated letter dated 11/28/2017. The letter informs the customer this letter replaces the instructions in the original notification dated 9/28/2017.
Quantity in commerce
268 units
Distribution
Worldwide Distribution - US Distribution to the states of : AZ, CA, CO, FL, GA, IL, IN, KS, NC, NY, OR, PA, SC, TX, and WA, including Puerto Rico and government distribution, and to the countries of : Canada, Angola, Australia, Belgium, Botswana, Brazil, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Hong Kong, India, Israel, Italy, Kenya, Lesotho, Malawi, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Norway, Pakistan, Panama, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, Uganda, Ukraine, United Kingdom, and Zambia.
Product codes and lots
All existing software versions (2.0, 2.0.1, and 2.1).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0091-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0091-2018Class IITerminatedVoluntary: Firm initiated