Z-0187-2018Class I

LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are...

Physio Control, Inc. / initiated 2017-10-27 / Terminated / root cause: manufacturing process

Physio-Control, Inc. began a recall of LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK... on . FDA classified it as Class I and gives the reason as "The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0187-2018
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Physio Control, Inc.
Product
LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm Physio Control, initiated their field action on 10/27/2017 by issuing a press release. The firm followed this with a recall notice dated 10/23/2017, to the direct consignees on 11/03/2017. The letter described the product, explained the problem and provided the following instructions: Quarantine the affected lots;contact Physio-Control at 1-866-231-1220, 6:00 A.M. to 4:00 P.M. (Pacific), Monday Friday to arrange for replacement of your unused electrodes and "Upon receipt of this notification, post a copy of the enclosed Correct Electrode and Packaging Labeling with each of your AEDs, which shows the correct placement of the electrodes." Physio_Control will provide replacement products for all unused affected Electrodes. Should you have any questions about this Product Recall, please contact us at 1-866-231-1220, 6:00 A.M. to 4:00 P.M. (Pacific), Monday Friday.
Quantity in commerce
14,219 units in total
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico and Guam and countries of: Canada, Austria, Belgium, Czech Republic; Denmark, Finland, France, French Polynesia, Germany, Hungary, Ireland, Italy, Latvia, Netherlands, Australia, China, Hong Kong, Japan, Argentina, Mexico, Norway, Panama, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, and New Zealand.
Product codes and lots
45932237, 46042286, 45979590, 46050960, 45979954, 46052545, 46007867, 46061770, 46023185, 46063054, 46023823, 46078012,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0187-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0186-2018Class ITerminatedVoluntary: Firm initiated
Z-0187-2018Class ITerminatedVoluntary: Firm initiated