Z-0075-2018Class II

HARVEST TERUMOBCT, GCP-10 Graft Delivery Pack, REF 51449, Rx Only, STERILE EO Harvest Graph Delivery System...

Terumo BCT, Ltd / initiated 2017-10-20 / Terminated / root cause: packaging

Terumo BCT, Inc. began a recall of HARVEST TERUMOBCT, GCP-10 Graft Delivery Pack, REF 51449, Rx Only, STERILE EO Harvest Graph Delivery System is a disposable pack that takes the cell... on . FDA classified it as Class II and gives the reason as "Certain lots of the Harvest Graft Delivery System GDP-10 Procedure Packs may have a lack of a seal on the inner pouch". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0075-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo BCT, Ltd
Product
HARVEST TERUMOBCT, GCP-10 Graft Delivery Pack, REF 51449, Rx Only, STERILE EO Harvest Graph Delivery System is a disposable pack that takes the cell concentrate created while using the BMAC Procedure Packs and allows for hydration of bone graft material for application to the sites doctors deem necessary
Product code
FMF Syringe, Piston
Reason for recall
Certain lots of the Harvest Graft Delivery System GDP-10 Procedure Packs may have a lack of a seal on the inner pouch.
Root cause tag
packaging
FDA root cause
Packaging
Action
Harvest Terumo sent a Voluntary Medical Device Product Recall letter dated October 20, 2017, to all affected customers. The letter stated the following: "ACTIONS REQUIRED FOR HEALTHCARE PROVIDERS AND DISTRIBUTORS 1. For product shelved in the original case (white box), or kit components, examine the outer label for catalog/lot numbers as shown in Figure A above. 2. Please return any unused affected product in your inventory. Please contact your local Terumo BCT Customer Support Center at 1.877.3.FYI.BCT (U.S. toll-free 1.877.339.4228) or +1.303.231.HELP (+1.303.231.4357), Terumo BCT Europe N.V. at +32.2.715.0590 or your local Terumo BCT representative, and Terumo BCT will issue a return goods authorization for product return. 3. Continue to use unaffected Harvest Graft Delivery System GDP-10 Procedure Packs in accordance with the instructions for use. 4. Distribute this notification to all Harvest Graft Delivery System GDP-10 Procedure Pack users within your organization. 5. IMPORTANT: Complete the attached acknowledgement form and return it by fax or email to Harvest Terumo BCT by 15 November 2017. Your return of the acknowledgement form is critical so we can confirm that you have received the recall notice. 6. As a reminder, please comply with the recommended practices for maintaining a sterile field as recommended by the Association of periOperative Registered Nurses (AORN) and refer to the instructions for use." Customers with questions, please contact your Terumo BCT representative or your regional Customer Support Center: U.S. Toll-Free: 1.877.3.FYI BCT (394 228) U.S.: +1.303.231.HELP (4357) Canada Toll-Free: 1.877.722.8411 Europe: +32.2.715.0590
Quantity in commerce
525 kits
Distribution
Nationwide Distribution including CA, CO, GA, LA, NJ, NV, NY, OR, PA, TX, and UT.
Product codes and lots
UDI Case 35020583514493, Each 05020583514492  Batch/Number: 04A9925, Exp. 01MAR2018; 05A9928, Exp. 01MAY2019; 06A9948, Exp. 01JUN2019; 06A9978, Exp. 01JUN2019; 08A9950, Exp. 01AUG2019

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0075-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0075-2018Class IITerminatedVoluntary: Firm initiated