Z-0101-2018Class II
Plum 360 Infusion System, List number 30010
Abbott Medical / initiated 2017-10-30 / Terminated
ICU Medical Inc began a recall of Plum 360 Infusion System, List number 30010 on . FDA classified it as Class II and gives the reason as "(1) Under certain conditions, if a malfunction alarm occurs while the pump is in the "Paused" state waiting for the distal pressure to decrease, the pump cannot be turned off and delivery cannot be restarted until the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0101-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Plum 360 Infusion System, List number 30010.
- Product code
- FRN Pump, Infusion
- Reason for recall
- (1) Under certain conditions, if a malfunction alarm occurs while the pump is in the "Paused" state waiting for the distal pressure to decrease, the pump cannot be turned off and delivery cannot be restarted until the battery is discharged or is disconnected; and (2) when the user accesses the Preventive Maintenance Screen in Service/Biomed Mode with a Total Delivery Time >1,500 hours, the user will not be able to interact with the device and the device must be power cycled.
- FDA root cause
- Under Investigation by firm
- Action
- The recalling firm issued letters dated 10/30/2017 via UPS on 10/30/2017 notifying their customers of the two issues.
- Quantity in commerce
- 21,461 devices
- Distribution
- Distribution was nationwide to medical facilities. Foreign distribution was made to Canada and Australia. Government and military distribution was also made.
- Product codes and lots
- All devices with software v15.10.00.010