Z-0101-2018Class II

Plum 360 Infusion System, List number 30010

Abbott Medical / initiated 2017-10-30 / Terminated

ICU Medical Inc began a recall of Plum 360 Infusion System, List number 30010 on . FDA classified it as Class II and gives the reason as "(1) Under certain conditions, if a malfunction alarm occurs while the pump is in the "Paused" state waiting for the distal pressure to decrease, the pump cannot be turned off and delivery cannot be restarted until the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0101-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Plum 360 Infusion System, List number 30010.
Product code
FRN Pump, Infusion
Reason for recall
(1) Under certain conditions, if a malfunction alarm occurs while the pump is in the "Paused" state waiting for the distal pressure to decrease, the pump cannot be turned off and delivery cannot be restarted until the battery is discharged or is disconnected; and (2) when the user accesses the Preventive Maintenance Screen in Service/Biomed Mode with a Total Delivery Time >1,500 hours, the user will not be able to interact with the device and the device must be power cycled.
FDA root cause
Under Investigation by firm
Action
The recalling firm issued letters dated 10/30/2017 via UPS on 10/30/2017 notifying their customers of the two issues.
Quantity in commerce
21,461 devices
Distribution
Distribution was nationwide to medical facilities. Foreign distribution was made to Canada and Australia. Government and military distribution was also made.
Product codes and lots
All devices with software v15.10.00.010

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0101-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0101-2018Class IITerminatedVoluntary: Firm initiated