Z-0282-2018Class II

Zimmer Dental Instrument Kit System, Tapered Screw-Vent Implant System surgical kit, Complete, Ref TSVKIT...

ZimVie US Corp LLC / initiated 2017-10-11 / Terminated / root cause: manufacturing process

Zimmer Dental Inc began a recall of Zimmer Dental Instrument Kit System, Tapered Screw-Vent Implant System surgical kit, Complete, Ref TSVKIT. The label shows the product is Manufactured by... on . FDA classified it as Class II and gives the reason as "The device may contain a manufacturing condition affecting the geometry of the driver's hex feature which prevents it from properly engaging the fixture mount transfer and/or a dental implant". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0282-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
Zimmer Dental Instrument Kit System, Tapered Screw-Vent Implant System surgical kit, Complete, Ref TSVKIT. The label shows the product is Manufactured by Zimmer Dental Inc., Carlsbad, CA. Endosseous dental implant accessories are manually powered devices intended to aid in the placement or removal of endosseous dental implants and abutments, prepare the site for placement of endosseous dental implants or abutments, aid in the fitting of endosseous dental implants or abutments, aid in the fabrication of dental prosthetics, and be used as an accessory with endosseous dental implants when tissue contact will last less than 1 hour.
Product code
NDP Accessories, Implant, Dental, Endosseous
Reason for recall
The device may contain a manufacturing condition affecting the geometry of the driver's hex feature which prevents it from properly engaging the fixture mount transfer and/or a dental implant.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The recalling firm, Zimmer Biomet, issued "URGENT MEDICAL DEVICE RECALL" letters addressed to Clinicians dated 10/11/2017 via FedEx on 10/11/2017. Letters to the distributors dated 10/11/2017 were issued via courier or email on 10/11/2017. The letters described the product, problem and actions to be taken. The customers were instructed to review notification for awareness; review your inventory; immediately return all affected product from your facility to: Field Action PM Regulatory Compliance Zimmer Biomet 4555 Riverside Dr. Palm Beach Gardens, FL 33410 US and complete and return the Certificate of Acknowledgment by fax to 574-372-4265 or email: [email protected]. If you have further questions or concerns after reviewing this notice, please call the Recall Department at 561-776-6700 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a prompt to leave a voicemail. Alternatively, your questions may be emailed to [email protected].
Quantity in commerce
891 kits
Distribution
Worldwide Distribution: US (nationwide) and countries of: Canada, Australia, Belgium, Bulgaria, Chile, Costa Rica, Egypt, France, Georgia, Germany, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Japan, Lebanon, Lithuania, Netherlands, Nicaragua, Poland, Portugal, Romania, Serbia, Shanghai, Spain, Taiwan, Thailand, Tunisia, Turkey, UK, and United Arab Emirates.
Product codes and lots
Lot numbers 63581758, 63583277, 63587717, 63623845, 63632463, 63632464, 63666560, 63669280, 63676567, 63676568, 63693941, 63694051, 63715410, 63716081, 63724569, 63724570, 63728840, 63735508, 63735987, 63739443, 63743229, 63748227, and 63750186

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0282-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0279-2018Class IITerminatedVoluntary: Firm initiated
Z-0280-2018Class IITerminatedVoluntary: Firm initiated
Z-0281-2018Class IITerminatedVoluntary: Firm initiated
Z-0282-2018Class IITerminatedVoluntary: Firm initiated