3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)
3M Company - Health Care Business began a recall of 3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000) on . FDA classified it as Class II and gives the reason as "During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0114-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- 3M Company
- Product
- 3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)
- Product code
- FLL Continuous Measurement Thermometer
- Reason for recall
- During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated their recall on 10/17/2017 via mail. International subsidiaries were notified via email, also on October 17, 2017.Distributors were asked to notify their customers and all direct consignees were directed to dispose of the product on site.
- Quantity in commerce
- 153 cases (3825 sensors)
- Distribution
- CA, MI, IL, MO, MN, GA, and Canada, Switzerland
- Product codes and lots
- Lot Numbers: 17230, Exp. 2020 Aug 18; 17233, Exp. 2020 Aug 21; 17234, Exp. 2020 Aug 22; 17241, Exp. 2020 Aug 29
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0114-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.