Z-0072-2018Class I

Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G...

GE Healthcare / initiated 2017-10-11 / Terminated / root cause: software

Spacelabs Healthcare, Ltd. began a recall of Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is... on . FDA classified it as Class I and gives the reason as "Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0072-2018
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
Root cause tag
software
FDA root cause
Software design
Action
Spacelabs Healthcare sent an Urgent Medical Device Correction letter dated October 24, 2017. The letter identified the affected product problem and actions to be taken. For questions call 1-800-522-7025 and select 4 for Technical Support.
Quantity in commerce
110 units
Distribution
Nationwide Distribution to the states of NM, WY, FL, NC, MS, CO, CT, ME, and AL.
Product codes and lots
UDI: 10841522100017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0072-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0072-2018Class ITerminatedVoluntary: Firm initiated