Z-0489-2018Class I

MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969

Philips North America LLC / initiated 2017-07-12 / Terminated / root cause: manufacturing process

Philips North America, LLC began a recall of MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular... on . FDA classified it as Class I and gives the reason as "Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0489-2018
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Philips Healthcare sent an Urgent Medical Device Correction Notification/Field Safety Notice letter dated June, 2017 to their customers. The letter identified the affected product, problem, hazard involved and actions to be taken. Philips is requesting that customers report if their device fails to function as intended. Philips will contact all customers for device replacement once parts are available. If you need further information or support concerning this action, please contact your local Philips representative.
Quantity in commerce
71 units
Distribution
No U.S. distribution, Internationally to: India, Ireland, Japan, Republic of Korea, Malaysia, Mexico, South Africa & United Kingdom
Product codes and lots
Serial Numbers:  US00601802, US00601968, US00601969, US00602114, US00602122, US00602124, US00602125, US00602126, US00602127, US00602128, US00602145, US00602180, US00602182, US00602184, US00602185, US00602186, US00602187, US00602189, US00602206, US00602236, US00602237, US00602238, US00602239, US00602240, US00602241, US00602242, US00602243, US00602244, US00602245, US00602246, US00602247, US00602248, US00602250, US00602251, US00602252, US00602253, US00602254, US00602344, US00602345, US00602346, US00602347, US00602348, US00602349, US00602350, US00602351, US00602352, US00602353, US00602354, US00602355, US00602356, US00602357, US00602358, US00602359, US00602360, US00602361, US00602362, US00602363, US00602364, US00602365, US00602366, US00602367, US00602368, US00602369, US00602370, US00602371, US00602372, US00602373, US00602374, US00602375, US00602376, US00602377

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0489-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0489-2018Class ITerminatedVoluntary: Firm initiated