MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969
Philips North America, LLC began a recall of MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular... on . FDA classified it as Class I and gives the reason as "Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0489-2018
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Philips Healthcare sent an Urgent Medical Device Correction Notification/Field Safety Notice letter dated June, 2017 to their customers. The letter identified the affected product, problem, hazard involved and actions to be taken. Philips is requesting that customers report if their device fails to function as intended. Philips will contact all customers for device replacement once parts are available. If you need further information or support concerning this action, please contact your local Philips representative.
- Quantity in commerce
- 71 units
- Distribution
- No U.S. distribution, Internationally to: India, Ireland, Japan, Republic of Korea, Malaysia, Mexico, South Africa & United Kingdom
- Product codes and lots
- Serial Numbers: US00601802, US00601968, US00601969, US00602114, US00602122, US00602124, US00602125, US00602126, US00602127, US00602128, US00602145, US00602180, US00602182, US00602184, US00602185, US00602186, US00602187, US00602189, US00602206, US00602236, US00602237, US00602238, US00602239, US00602240, US00602241, US00602242, US00602243, US00602244, US00602245, US00602246, US00602247, US00602248, US00602250, US00602251, US00602252, US00602253, US00602254, US00602344, US00602345, US00602346, US00602347, US00602348, US00602349, US00602350, US00602351, US00602352, US00602353, US00602354, US00602355, US00602356, US00602357, US00602358, US00602359, US00602360, US00602361, US00602362, US00602363, US00602364, US00602365, US00602366, US00602367, US00602368, US00602369, US00602370, US00602371, US00602372, US00602373, US00602374, US00602375, US00602376, US00602377
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0489-2018 | Class I | Terminated | Voluntary: Firm initiated |