Z-0184-2018Class II

ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is...

Zimmer GmbH / initiated 2014-09-05 / Terminated / root cause: software

Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial... on . FDA classified it as Class II and gives the reason as "The software issue described was corrected in the modification to the MXTTOUT controller parameter settings". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0184-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
The software issue described was corrected in the modification to the MXTTOUT controller parameter settings.
Root cause tag
software
FDA root cause
Software design
Action
This is a retrospective report of a correction initiated on 05-Sept-2014. The software issue described was corrected in the modification to the MXTTOUT controller parameter settings. There were 8 affected devices at customer sites in the US. Field Service Technicians were deployed to the consignee locations to perform the system upgrade.
Quantity in commerce
29 (8 US and 21 OUS)
Distribution
Nationwide Distribution to AK, OH, TX, GA, and MI
Product codes and lots
Serial #'s: RO10009, RO10011, RO10014, RO13023, RO13027, RO14031, RO14033 and RO14035

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0184-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0184-2018Class IITerminatedVoluntary: Firm initiated