ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument...
Zimmer Biomet, Inc. began a recall of ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to... on . FDA classified it as Class II and gives the reason as "Communication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0167-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Communication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- This is a retrospective report of a correction initiated on 20-Jul-2016. Customers were informed onsite by field Service Technicians of the planned correction. The software issue described was corrected in the new software version ROSA Brain 3.0.0.20. Field Service Technicians were deployed to the customers locations to perform the system upgrade. .
- Quantity in commerce
- 18 (14 US and 4 OUS)
- Distribution
- Worldwide Distribution - US Nationwide in the states of OH, FL, MA, MN, DC, NC, CA, NY, PA and countries of Australia and France
- Product codes and lots
- Serial #'s: BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, and BR16021