Z-0167-2018Class II

ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument...

Zimmer GmbH / initiated 2016-07-20 / Terminated / root cause: software

Zimmer Biomet, Inc. began a recall of ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to... on . FDA classified it as Class II and gives the reason as "Communication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0167-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Communication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller.
Root cause tag
software
FDA root cause
Software design
Action
This is a retrospective report of a correction initiated on 20-Jul-2016. Customers were informed onsite by field Service Technicians of the planned correction. The software issue described was corrected in the new software version ROSA Brain 3.0.0.20. Field Service Technicians were deployed to the customers locations to perform the system upgrade. .
Quantity in commerce
18 (14 US and 4 OUS)
Distribution
Worldwide Distribution - US Nationwide in the states of OH, FL, MA, MN, DC, NC, CA, NY, PA and countries of Australia and France
Product codes and lots
Serial #'s: BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, and  BR16021

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0167-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0167-2018Class IITerminatedVoluntary: Firm initiated