Z-0102-2018Class II

ROSA Surgical Device 2.5.8. It is intended to be used in the operating room for the spatial positioning and...

Zimmer GmbH / initiated 2014-02-14 / Terminated / root cause: labeling and UDI

Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide on . FDA classified it as Class II and gives the reason as "An undetected shift between the information displayed in the navigation software and the actual patient anatomy". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0102-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ROSA Surgical Device 2.5.8. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
An undetected shift between the information displayed in the navigation software and the actual patient anatomy
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
There were 4 units imported into the US and all have been corrected.
Quantity in commerce
21 (4 US and 17 OUS)
Distribution
OH, MI and TX
Product codes and lots
Serial #s; RO10009, RO10011, RO10014, and RO13023

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0102-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0102-2018Class IITerminatedVoluntary: Firm initiated