ROSA Surgical Device 2.5.8. It is intended to be used in the operating room for the spatial positioning and...
Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide on . FDA classified it as Class II and gives the reason as "An undetected shift between the information displayed in the navigation software and the actual patient anatomy". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0102-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ROSA Surgical Device 2.5.8. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- An undetected shift between the information displayed in the navigation software and the actual patient anatomy
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- There were 4 units imported into the US and all have been corrected.
- Quantity in commerce
- 21 (4 US and 17 OUS)
- Distribution
- OH, MI and TX
- Product codes and lots
- Serial #s; RO10009, RO10011, RO10014, and RO13023
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0102-2018 | Class II | Terminated | Voluntary: Firm initiated |