NucliSENS Lysis Buffer Extended lot for RES 76675
BioMerieux SA began a recall of NucliSENS Lysis Buffer Extended lot for RES 76675 on . FDA classified it as Class II and gives the reason as "Elute may become colored due to the residual presence of heme group origination from hemoglobin present in whole blood samples as well as dried spot (DBS) samples. This can potentially cause the inhibition of PCR...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0086-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- NucliSENS Lysis Buffer Extended lot for RES 76675
- Product code
- PPM General Purpose Reagent
- Reason for recall
- Elute may become colored due to the residual presence of heme group origination from hemoglobin present in whole blood samples as well as dried spot (DBS) samples. This can potentially cause the inhibition of PCR reactions and result in an uninterpretable test result. If using an Internal Control (IC) in the extraction process the IC could potentially be inhibited as well and invalidate test results.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- US customers were notified on 10/18/2017 via letter using 2-day Fed Ex shipment. Instructions included notifying all appropriate personnel in the laboratory or to customers if product was further distributed, refrain from using the affected product, and arrange for replacement of affected product by completing and returning the response form. It also instructed customers to contact and discuss any concerns regarding previously reported results with the appropriate Laboratory Medical Director to determine an appropriate course of action.
- Quantity in commerce
- 911 units
- Distribution
- Distributed domestically to MD, CO, and OH.
- Product codes and lots
- REF 200292, Lot 17022802, Exp Date 01/28/2019 Extended lot for RES 76675
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0086-2018 | Class II | Terminated | Voluntary: Firm initiated |