PICC Tray, STERILE, 12 packs each per case
Medline Industries, Inc. began a recall of PICC Tray, STERILE, 12 packs each per case on . FDA classified it as Class II and gives the reason as "Multiple surgical packs were assembled with a non-sterile CHG disk labeled as BIOPATCH" (kit component #52655) attached to the outside of the sterile pack. This may pose potential risk of infection. However, the kits...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2977-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- PICC Tray, STERILE, 12 packs each per case.
- Product code
- N/A
- Reason for recall
- Multiple surgical packs were assembled with a non-sterile CHG disk labeled as BIOPATCH" (kit component #52655) attached to the outside of the sterile pack. This may pose potential risk of infection. However, the kits have been in distribution since March 2016. Many of these kits have been used with no reported infections or adverse patient events. In addition, the BIOPATCH is impregnated with CHG, a well-known antiseptic agent with broad-spectrum anti-microbial and antifungal activity.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Firm contacted customers via phone and letter beginning April 12, 2017. Customers were instructed to remove the Ethicon BioPatch Protective Disk with CHG from the kit(s) and return it to the firm. The firm would provide a sterile replacement BioPatch. Affected product will be destroyed.
- Quantity in commerce
- 144 kits
- Distribution
- CA, NM
- Product codes and lots
- Item number: DYNJ49177C. Lots: 16HB4700, 16UB5076, 17MB6837
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2976-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2977-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2978-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2979-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2980-2018 | Class II | Terminated | Voluntary: Firm initiated |