Z-2978-2018Class II
Laparoscopy CDS
MEDLINE INDUSTRIES, LP - Northfield / initiated 2017-04-12 / Terminated / root cause: labeling and UDI
Medline Industries, Inc. began a recall of Laparoscopy CDS on . FDA classified it as Class II and gives the reason as "One manufacturing lot of a non-sterile pack was assembled with a non-sterile component labeled sterile (kit component #80532NS). The non-sterile component 80532NS was supposed to be placed inside the sterile portion of...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2978-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Laparoscopy CDS
- Product code
- N/A
- Reason for recall
- One manufacturing lot of a non-sterile pack was assembled with a non-sterile component labeled sterile (kit component #80532NS). The non-sterile component 80532NS was supposed to be placed inside the sterile portion of the pack where it would subsequently be exposed to the ethylene oxide sterilization process. However, the component was not placed in the sterile portion of the pack during assembly and was never exposed to the sterilization process. Component 80532NS was inadvertently released non-sterile. Component 80532NS is packaged by Medline and is an Anti-fog kit that contains 1 bottle of 6g Anti-Fog solution and an adhesive backed sponge.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Firm contacted customers via phone and letter beginning April 12, 2017. Customers were instructed to remove the Ethicon BioPatch Protective Disk with CHG from the kit(s) and return it to the firm. The firm would provide a sterile replacement BioPatch. Affected product will be destroyed.
- Quantity in commerce
- 15 kits
- Distribution
- CA, NM
- Product codes and lots
- Model: CDS920114F. Lot: 17MB5499
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2976-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2977-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2978-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2979-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2980-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY