510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2978-2018Class II

Laparoscopy CDS

MEDLINE INDUSTRIES, LP - Northfield / initiated 2017-04-12 / Terminated / root cause: labeling and UDI

Medline Industries, Inc. began a recall of Laparoscopy CDS on . FDA classified it as Class II and gives the reason as "One manufacturing lot of a non-sterile pack was assembled with a non-sterile component labeled sterile (kit component #80532NS). The non-sterile component 80532NS was supposed to be placed inside the sterile portion of...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2978-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Laparoscopy CDS
Product code
N/A
Reason for recall
One manufacturing lot of a non-sterile pack was assembled with a non-sterile component labeled sterile (kit component #80532NS). The non-sterile component 80532NS was supposed to be placed inside the sterile portion of the pack where it would subsequently be exposed to the ethylene oxide sterilization process. However, the component was not placed in the sterile portion of the pack during assembly and was never exposed to the sterilization process. Component 80532NS was inadvertently released non-sterile. Component 80532NS is packaged by Medline and is an Anti-fog kit that contains 1 bottle of 6g Anti-Fog solution and an adhesive backed sponge.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Firm contacted customers via phone and letter beginning April 12, 2017. Customers were instructed to remove the Ethicon BioPatch Protective Disk with CHG from the kit(s) and return it to the firm. The firm would provide a sterile replacement BioPatch. Affected product will be destroyed.
Quantity in commerce
15 kits
Distribution
CA, NM
Product codes and lots
Model: CDS920114F. Lot: 17MB5499

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2978-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2976-2018Class IITerminatedVoluntary: Firm initiated
Z-2977-2018Class IITerminatedVoluntary: Firm initiated
Z-2978-2018Class IITerminatedVoluntary: Firm initiated
Z-2979-2018Class IITerminatedVoluntary: Firm initiated
Z-2980-2018Class IITerminatedVoluntary: Firm initiated