Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0523-2018Class II

Disposable Light Handle Cover, Item Number B1-715-65. Packaged 1 device inside a peel pouch and sealed...

Microtek Medical LLC / initiated 2017-10-13 / Terminated / root cause: packaging

Ecolab, Inc. began a recall of Disposable Light Handle Cover, Item Number B1-715-65. Packaged 1 device inside a peel pouch and sealed. Sealed pouches are packaged 10 each inside a poly... on . FDA classified it as Class II and gives the reason as "Sterile packaging of the disposable light handle cover may contain gaps in the seal of the pouch". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0523-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microtek Medical LLC
Product
Disposable Light Handle Cover, Item Number B1-715-65. Packaged 1 device inside a peel pouch and sealed. Sealed pouches are packaged 10 each inside a poly liner. 5 rows of 10 pouches are placed in a labeled corrugated mailer box. Product Usage: Intended to be used as a disposable barrier for a surgical lighting system and which allows a surgical team member to manually adjust the lighting system. Single use device.
Product code
FTA Light, Surgical, Accessories
Reason for recall
Sterile packaging of the disposable light handle cover may contain gaps in the seal of the pouch.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
Ecolab sent and Urgent Medical Device Recall letter dated October 13, 2017 to customers. The identified the affected product, problem and actions to be taken. Instructions include to immediately quarantine lots in inventory, notify customers since the product has been further distributed, return affected product, and complete and return the Recall Response Form. For questions call 903-597-2568.
Quantity in commerce
5,958 cases (50 per case)
Distribution
Distributed domestically to Michigan.
Product codes and lots
Model/Item/Part No. B1-715-65, Lot Numbers: 14224, 14225, 14226, 14237, 14238, 14239, 14240, 14255, 14253, 14254, 14252, 14261, 14266, 14260, 14265, 14262, 14279, 14276, 14269, 14275, 14273, 14258, 14274, 14281, 14286, 14287, 14283, 14282, 14288, 14272, 14280, 14289, 14308, 14307, 14309, 14310, 14311, 14337, 14342, 14335, 14338, 14345, 14346, 14336,  14343, 14339, 14344, 15015, 15013, 15019, 15020, 15021, 15022, 15030, 15029, 15028, 15033, 15036, 15037, 15035, 15034, 15041, 15043, 15042, 15044, 15055, 15048, 15056, 15054, 15083, 15078, 15072, 15082, 15070, 15068, 15079, 15142, 15146, 15154, 15149, 15148, 15153, 15147, 15084, 15047, 15049, 15085, 15026, 15086, 15012, 15051, 15009, 15027, 15091, 15090, 15092, 15096, 15050, 15098, 15089, 15097, 15100, 15216, 15218, 15103, 15224, 15212, 15211, 15215, 15222, 15231, 15239, 15240, 15232, 15229, 15230, 15243, 15226, 15237, 15252, 15261, 15266, 15247, 15251, 15264, 15265, 15272, 15275, 15274, 15268, 15267, 15279, 15271, 15280, 15286, 15282, 15296, 15295, 15300, 15301, 15306, 15294, 15302, 15308, 15313, 15315, 15317, 15322, 15321, 15316, 15323, 15328, 15336, 15334, 15327, 15324, 15337, 15342, 15357, 16006, 16008, 15351, 15352, 16007, 15341, 16071, 15343, 15344, 16076, 16078, 16089, 16081, 16082, 16083, 16091, 16092, 16109, 16103, 16106, 16110, 16099, 16098, 16116, 16134, 16139, 16144, 16137, 16140, 16141, 16145, 16172, 16174, 16176, 16173, 16175, 16168, 16190, 16169, 16193, 16194, 16195, 16196, 16217, 16221, 16222, 16223, 16229, 16224, 16225, 16228, 16245, 16232, 16230, 16235, 16236, 16237, 16242, 16251, 16252, 16208, 16210, 16209, 16214, 16211, 16257, 16253, 16256, 16258, 16270, 16271, 16272, 16273, 16277, 16349, 16350, 16354, 16351, 16355, 16356, 17002, 16357, 17011, 17004, 17009, 17012, 17040, 17013, 17041, 17039, 17045, 17046, 17044, 17052, 17053, 17058, 17054, 17060, 17061, 17065, 17066, 17081, 17086, 17089, 17093, 17094, 17090, 17095, 17088, 17097, 17177, 17100, 17186, 17180, 17188, 17192, 17191

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0523-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0523-2018Class IITerminatedVoluntary: Firm initiated