Rosa Spine 1.0.2 Stereotaxic Instrument Computer-Assisted Surgical Device The device is intended for the...
Zimmer Biomet, Inc. began a recall of Rosa Spine 1.0.2 Stereotaxic Instrument Computer-Assisted Surgical Device The device is intended for the spatial positioning and orientation of instrument... on . FDA classified it as Class II and gives the reason as "A design change was initiated to update ROSA Spine 1.0.2 to version ROSA Spine 1.0.2.16 to resolve software bugs and improve usability and stability of the ROSA Spine device". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0177-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Rosa Spine 1.0.2 Stereotaxic Instrument Computer-Assisted Surgical Device The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by surgeons to guide standard neurosurgical instruments during spine surgery. Guidance is based on an intra-operative plan developed with three-dimensional imaging software provided that the required fiducial markers and rigid patient anatomy can be identified on 3D CT scans. The device is indicated for the placement of pedicle screws in lumber vertebrae with a posterior approach.
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- A design change was initiated to update ROSA Spine 1.0.2 to version ROSA Spine 1.0.2.16 to resolve software bugs and improve usability and stability of the ROSA Spine device.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Zimmer Biomet issued a retrospective report of a correction taken on 29-Oct-2015. The software issue described was corrected in the new software version 1.0.2.16. There were 2 copies of the affected software in distribution. Field Service Technicians were deployed to the consignee locations to perform the system upgrade. 100% of affected devices were updated with the software solution. No further action is required.
- Quantity in commerce
- 2
- Distribution
- Worldwide Distribution including France and Germany
- Product codes and lots
- Serial #'s; SP14003 and SP14005
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0177-2018 | Class II | Terminated | Voluntary: Firm initiated |