nordicBrainEx 2.0 provides analysis and visualization capabilities of dynamic MRI data of the brain...
NordicNeuroLab AS began a recall of nordicBrainEx 2.0 provides analysis and visualization capabilities of dynamic MRI data of the brain, presenting the derived properties and parameters in a... on . FDA classified it as Class II and gives the reason as "An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2050-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- NordicNeuroLab AS
- Product
- nordicBrainEx 2.0 provides analysis and visualization capabilities of dynamic MRI data of the brain, presenting the derived properties and parameters in a clinically useful context.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm notified their consignees by email on 09/17/2014.
- Quantity in commerce
- 22 licenses
- Distribution
- USA, Sweden, Spain, Russia, Hungary, India, Norway, Iceland, Italy, South Korea, United Kingdom
- Product codes and lots
- version 2.1 and older
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2050-2018 | Class II | Terminated | Voluntary: Firm initiated |