510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0256-2018Class II

TERUMO(R) Overpressure Safety Valve, REF: (a) LH130 (b) LH130J

Terumo Cardiovascular Systems Corporation / initiated 2017-06-27 / Terminated

Terumo Cardiovascular Systems Corporation began a recall of TERUMO(R) Overpressure Safety Valve, REF: (a) LH130 (b) LH130J on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems (Terumo CVS) has received reports of no flow through the over-pressure safety (OPS) Valve". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0256-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
TERUMO(R) Overpressure Safety Valve, REF: (a) LH130 (b) LH130J
Product code
DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) has received reports of no flow through the over-pressure safety (OPS) Valve.
FDA root cause
Process change control
Action
The firm sent a safety alert to its consignees on 07/27/2017 by mail. The notice requested that the consignee test the product prior to using. If the product did not operate, the consignee was asked to return it. Beginning on 12/11/2017 the firm conducted a removal by mail of any remaining product. The customer instructions are as follows: CUSTOMER INSTRUCTIONS Review this Medical Device Recall Notice. Assure that all users receive notice of this issue. Refer to the Customer Response Form to identify your product that is subject to this action. Confirm receipt of this notification by completing and returning the attached Customer Response Form to the email address or fax number indicated on the form . Terumo CVS will issue a Returned Goods Authorization upon receipt of the Customer Response Form. Return all affected products to Terumo CVS. QUESTIONS? We encourage you to contact Terumo CVS with any questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 Monday- Friday, 8 a.m . - 6 p.m. ET Terumo Safety Alert Fax: 1.410.392.7183 Terumo Safety Alert Email: [email protected]
Quantity in commerce
297,222 units
Distribution
Worldwide distribution: US, Canada, Colombia, UK, Belgium, and Japan
Product codes and lots
SAFETY ALERT Lot Codes:  (a) RH14, RK11, RK20, RL02, RM06, RM20, RN10, RP08, TD16, TE13, TE27, TF14, TF21, TG04, TG11, TG17, TH06, TH16, TK10, TK24, TL08, TL21, TM28, TP14, UA11, UA18, UA25, UC08, UC22, UD14, UD21, UD28, UE05, UE18, UF09, UG06, UK01, UL12, UM10, UM31, UP05, VA02, VC06, VD06, VE10    (b) RH14, RH21, RK18, RL15, RM01, RM22, RN24, RP01, TA19, TF05, TF27, TG19, TH20, TK24, TL01, TM12, TN09, TP03, UA20, UC15, UD22, UE04, UE18, UF02, UG13, UH11, UK15, UL06, UM10, VA02, VA30, VC27, VE03, VF01   LH130 PRODUCT CORRECTION lot codes or serial number range: TA19 through VD06, distribution date range: 08Apr2015 through 20Jan2017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0256-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0256-2018Class IITerminatedVoluntary: Firm initiated
Z-0257-2018Class IITerminatedVoluntary: Firm initiated