Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery...
Implant Direct Sybron Manufacturing, LLC began a recall of Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to... on . FDA classified it as Class II and gives the reason as "The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0283-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Implant Direct Sybron Manufacturing LLC
- Product
- Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consists of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.
- Product code
- NDP Accessories, Implant, Dental, Endosseous
- Reason for recall
- The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide was too large.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- On November 22, 2017, the recalling firm will send consignees customer letters with replacement product, requesting nonconforming product to be returned to Implant Direct with return labels and packaging. The recalling firm will then initiate calls 24 hours after the communication is sent to the clinician; initiate second calls 72 hours after the communication is sent; and initiate third calls 5 working days after the communication is sent. For further questions, please call (818) 444-3300.
- Quantity in commerce
- 560 units (worldwide)
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Europe, Canada, Australia, Egypt, Jordan, and Lebanon.
- Product codes and lots
- Lot # 38665, 38825, 42562, 42649, 43810, 45014, 46950, 50165, 45018, 50165, 52360, 52987, 60376, 62435, 64245, 67044, 73431, 73729, 74772, 70293
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0283-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.