Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0283-2018Class II

Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery...

Implant Direct Sybron Manufacturing LLC / initiated 2017-11-15 / Terminated / root cause: labeling and UDI

Implant Direct Sybron Manufacturing, LLC began a recall of Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to... on . FDA classified it as Class II and gives the reason as "The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0283-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Implant Direct Sybron Manufacturing LLC
Product
Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) Spectra System/Legacy Guided Surgery Handle Kit (Part Number: G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consists of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.
Product code
NDP Accessories, Implant, Dental, Endosseous
Reason for recall
The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide was too large.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
On November 22, 2017, the recalling firm will send consignees customer letters with replacement product, requesting nonconforming product to be returned to Implant Direct with return labels and packaging. The recalling firm will then initiate calls 24 hours after the communication is sent to the clinician; initiate second calls 72 hours after the communication is sent; and initiate third calls 5 working days after the communication is sent. For further questions, please call (818) 444-3300.
Quantity in commerce
560 units (worldwide)
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Europe, Canada, Australia, Egypt, Jordan, and Lebanon.
Product codes and lots
Lot # 38665, 38825, 42562, 42649, 43810, 45014, 46950, 50165, 45018, 50165, 52360, 52987, 60376, 62435, 64245, 67044, 73431, 73729, 74772, 70293

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0283-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0283-2018Class IITerminatedVoluntary: Firm initiated