Z-0263-2018Class II
Xper Flex Cardio Physiomonitoring system. Used to facilitate invasive investigations of heart and vascular...
Invivo Corporation / initiated 2017-08-18 / Terminated / root cause: software
Invivo Corporation began a recall of Xper Flex Cardio Physiomonitoring system. Used to facilitate invasive investigations of heart and vascular disease when non-invasive indicators warrant such on . FDA classified it as Class II and gives the reason as "Intermittent communication between the host system and the FlexCardio". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0263-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Invivo Corporation
- Product
- Xper Flex Cardio Physiomonitoring system. Used to facilitate invasive investigations of heart and vascular disease when non-invasive indicators warrant such.
- Product code
- MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
- Reason for recall
- Intermittent communication between the host system and the FlexCardio
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- On August 24, 2017 Philips send the following communications to distributors and clients. Dear Distributor, Philips is initiating an action related to the Xper Flex Cardio Patient Monitoring System. The attached Field Safety Notice is intended to inform customers about: " what the problem is and under what circumstances it can occur " the actions that should be taken by the customer / user in order to prevent risks for patients, users or servicers " the actions planned by Philips to correct the problem. It is imperative that all affected end users of the Xper Flex Cardio receive the attached Field Safety Notice. Because Philips sells these products through distributors, including your organization, we may not have the information necessary to contact all end users. Therefore, please send the attached Field Safety Notice to each customer to whom you have distributed one of the affected devices. A description of the affected device can be found in the Field Safety Notice attached. -- Dear Customer, A problem has been detected in the Philips Xper Flex Cardio Patient Monitoring System (Flex Cardio) that, if it were to recur, could pose a risk for patients or users. This Field Safety Notice is intended to inform you about: " what the problem is and under what circumstances it can occur " the actions that should be taken by the customer / user in order to prevent risks for patients or users " the actions planned by Philips to correct the problem. Because of intermittent communication between the host system and the Flex Cardio, it is possible for any of the following conditions to occur, which could result in delays in diagnosis or treatment: " Boom Monitor (display) may not display all active waveform and/or vital sign data. " Delayed audible and visual alarms at the Flex Cardio and Boom Monitor respectively, due to a mismatch between the alarm limits on the host system and the Flex Cardio/Boom Monitor. " After patient admission and ca
- Quantity in commerce
- 4,535
- Distribution
- AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Puerto Rico Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Bulgaria, Cambodia, Canada, Cayman Island, Chile, China, Colombia, Congo, Cote D'Ivoire, Cuba, Cyprus, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Honduras, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Korea, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Mauritius, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Papua, Paraguay, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, and Yemen
- Product codes and lots
- All Flex Cardio devices, Revision A, B Code Number's: 453564241901, 453564241911, 453564483321, 453564483331, 989803185601, 9898031930191, 860335 and 860338
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0263-2018 | Class II | Terminated | Voluntary: Firm initiated |