Z-0275-2018Class II

Stroke Fast Pack(TM), Trevo(TM) XP, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm; Excelsior(TM) XT-27(tm)...

Stryker Corp. / initiated 2017-11-03 / Terminated / root cause: labeling and UDI

Stryker Neurovascular began a recall of Stroke Fast Pack(TM), Trevo(TM) XP, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm; Excelsior(TM) XT-27(tm), MICROCATHETER,150 cm, 6 cm; AXS Catalyst(TM) 6, Distal... on . FDA classified it as Class II and gives the reason as "Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs and Trevo Procedure Packs were manufactured using a carton sleeve where the sleeve label contents did not match the physical contents within the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0275-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stroke Fast Pack(TM), Trevo(TM) XP, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm; Excelsior(TM) XT-27(tm), MICROCATHETER,150 cm, 6 cm; AXS Catalyst(TM) 6, Distal Access Catheter, 0.060 in, 132 cm, UPN M0033PK62523002 Stroke intervention kit
Product code
NRY Catheter, Thrombus Retriever
Reason for recall
Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs and Trevo Procedure Packs were manufactured using a carton sleeve where the sleeve label contents did not match the physical contents within the pack.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
The firm, Stryker Neurovascular, sent an "Urgent Medical Device Voluntary Recall Immediate Action Required" letter initiating their recall on 11/01/2017. The letter described the product, problem and actions to be taken. The consignees were instructed to immediately check internal inventory; remove and discard the procedure pack carton sleeve; circulate notice; maintain awareness of notice internally until all required actions have been completed with your facility; inform Stryker of any subject devices distributed to other organization, and complete and return customer response form to your nominated Stryker Representative or to [email protected]. If you have any questions, call 510-413-2593; email: [email protected] or 510-413-2900.
Quantity in commerce
2 units
Distribution
International Distribution to: Germany, Slovakia and Israel.
Product codes and lots
Lot Numbers: QXC10200044, exp. date 28-Aug-18; QXC10200043, exp. date 28-Aug-18

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0275-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0275-2018Class IITerminatedVoluntary: Firm initiated
Z-0276-2018Class IITerminatedVoluntary: Firm initiated
Z-0277-2018Class IITerminatedVoluntary: Firm initiated