Stroke Fast Pack(TM), Trevo(TM) XP, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm; Excelsior(TM) XT-27(tm)...
Stryker Neurovascular began a recall of Stroke Fast Pack(TM), Trevo(TM) XP, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm; Excelsior(TM) XT-27(tm), MICROCATHETER,150 cm, 6 cm; AXS Catalyst(TM) 6, Distal... on . FDA classified it as Class II and gives the reason as "Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs and Trevo Procedure Packs were manufactured using a carton sleeve where the sleeve label contents did not match the physical contents within the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0275-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stroke Fast Pack(TM), Trevo(TM) XP, TREVO(TM) XP PROVUE RETRIEVER, 6 mm, 25 mm; Excelsior(TM) XT-27(tm), MICROCATHETER,150 cm, 6 cm; AXS Catalyst(TM) 6, Distal Access Catheter, 0.060 in, 132 cm, UPN M0033PK62523002 Stroke intervention kit
- Product code
- NRY Catheter, Thrombus Retriever
- Reason for recall
- Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs and Trevo Procedure Packs were manufactured using a carton sleeve where the sleeve label contents did not match the physical contents within the pack.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- The firm, Stryker Neurovascular, sent an "Urgent Medical Device Voluntary Recall Immediate Action Required" letter initiating their recall on 11/01/2017. The letter described the product, problem and actions to be taken. The consignees were instructed to immediately check internal inventory; remove and discard the procedure pack carton sleeve; circulate notice; maintain awareness of notice internally until all required actions have been completed with your facility; inform Stryker of any subject devices distributed to other organization, and complete and return customer response form to your nominated Stryker Representative or to [email protected]. If you have any questions, call 510-413-2593; email: [email protected] or 510-413-2900.
- Quantity in commerce
- 2 units
- Distribution
- International Distribution to: Germany, Slovakia and Israel.
- Product codes and lots
- Lot Numbers: QXC10200044, exp. date 28-Aug-18; QXC10200043, exp. date 28-Aug-18
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0275-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0276-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0277-2018 | Class II | Terminated | Voluntary: Firm initiated |