Z-0237-2018Class II

UniCel DxH600, Part Number B23858, automated hematology analyzer (Note: the DxH 600 is the same instrument as...

Siemens Healthcare Diagnostics Inc. / initiated 2017-11-03 / Terminated / root cause: design

Beckman Coulter Inc. began a recall of UniCel DxH600, Part Number B23858, automated hematology analyzer (Note: the DxH 600 is the same instrument as the DxH 800 except that it is a bench top unit... on . FDA classified it as Class II and gives the reason as "The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0237-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
UniCel DxH600, Part Number B23858, automated hematology analyzer (Note: the DxH 600 is the same instrument as the DxH 800 except that it is a bench top unit with no instrument cart, computer stand, and no connectivity capabilities)
Product code
GKZ Counter, Differential Cell
Reason for recall
The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
Root cause tag
design
FDA root cause
Device Design
Action
The firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions: "To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed. You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance."
Quantity in commerce
1 unit
Distribution
CA, AR, HI, CA, OH, WA, MD, Canada, Japan
Product codes and lots
Serial Number AZ06682

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0237-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0236-2018Class IITerminatedVoluntary: Firm initiated
Z-0237-2018Class IITerminatedVoluntary: Firm initiated
Z-0238-2018Class IITerminatedVoluntary: Firm initiated