Z-0309-2018Class III

Access 2 Immunoassay System, Remanufactured, Catalog Number 386220. In vitro diagnostic device used for the...

Siemens Healthcare Diagnostics Inc. / initiated 2017-11-06 / Terminated / root cause: design

Beckman Coulter Inc. began a recall of Access 2 Immunoassay System, Remanufactured, Catalog Number 386220. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative... on . FDA classified it as Class III and gives the reason as "Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0309-2018
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access 2 Immunoassay System, Remanufactured, Catalog Number 386220. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.
Root cause tag
design
FDA root cause
Device Design
Action
An Urgent Medical Device Recall letter, dated November 6, 2017, were sent to the affected customers via mail. The letter identifies the affected device, the reason for recall, impact, and actions to be taken. The firm states that customers may continue to operate their system normally. Customers are instructed to contact Technical Support if a device on the Access 2 instrument needs alignment. The alignment tool is only to be used by the Beckman Coulter service personnel and Beckman Coulter trained biomedical engineers. A Beckman Coulter service rep will contact customers to schedule a service visit. Customers are to share this information with their staff and retain a copy of the letter. Customers are to respond via one of the ways identified. Beckman Coulter should be contacted with any questions regarding the notice.
Quantity in commerce
3 units
Distribution
Worldwide Distribution -- US, Argentina, Canada, Chile, China, Columbia, Costa Rica, France, Georgia, Germany, Ghana, Greece, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Morocco, Myanmar, Nepal, Philippines, Poland, Russia, Rwanda, Saudi Arabia, Slovakia, South Africa, South Korea, Thailand, Turkey, United Kingdom, and Vietnam.
Product codes and lots
Serial/Lot Numbers: 507114, 503299 and 508111

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0309-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0308-2018Class IIITerminatedVoluntary: Firm initiated
Z-0309-2018Class IIITerminatedVoluntary: Firm initiated
Z-0310-2018Class IIITerminatedVoluntary: Firm initiated