Tapered Screw-Vent MTX Model TSVTB11
ZimVie US Corp LLC / initiated 2017-11-20 / Terminated / root cause: manufacturing process
Zimmer Dental Inc began a recall of Tapered Screw-Vent MTX Model TSVTB11 on . FDA classified it as Class II and gives the reason as "Some implant vial caps in the recalled implant lot may inadvertently crack as the result of a manufacturing assembly condition". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0284-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZimVie US Corp LLC
- Product
- Tapered Screw-Vent MTX Model TSVTB11
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- Some implant vial caps in the recalled implant lot may inadvertently crack as the result of a manufacturing assembly condition.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- Recall notification letters will be sent to distributors and clinicians on or about the last week of November 2017. International distributors will be notified via courier or email. Clinicians will be instructed to; 1. Follow any specified patient monitoring instructions, 2. Return Attachment 1 (Acknowledgment of Responsibility) to Zimmer Biomet, 3. Review their inventory and returning any affected product.
- Quantity in commerce
- 500 units
- Distribution
- US, Canada, France, Italy, Spain, Germany
- Product codes and lots
- lot,#63519794; UDI: (01)00889024019935 (17)211130 (10)63519794
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0284-2018 | Class II | Terminated | Voluntary: Firm initiated |