Z-0346-2018Class I

Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number...

Jaeger Medical GmbH / initiated 2014-11-29 / Terminated / root cause: manufacturing process

Vyaire Medical began a recall of Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated... on . FDA classified it as Class I and gives the reason as "Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0346-2018
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Jaeger Medical GmbH
Product
Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.
Product code
BTT Humidifier, Respiratory Gas, (Direct Patient Interface)
Reason for recall
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm, Vyaire Medical, sent an "URGENT RECALL NOTIFICATION" letter dated November 29th, 2017 to its customers. The recall letter described the product, problem and actions to be taken. The customers were instructed to do the following: "ALL CUSTOMERS: ~ Inspect current inventory on-hand. A 100% physical inventory should immediately be performed to identify and remove affected lots of the AirLife Humidification Chamber(s) and Heated Breathing Circuit Kit(s) due to the identified potential patient safety risk. ~ Destroy all affected product(s) in-stock in accordance with your facility's destruction protocol. ~ If you are a healthcare facility or distributor that has further distributed affected product you must further notify your customers of this URGENT RECALL NOTIFICATION. CUSTOMERS THAT HAVE PURCHASED DIRECT FROM VYAIRE: ~ Complete the Customer Response Form (Appendix 2) and return to [email protected] ~ If you wish to receive credit of your in-stock product affected by this recall or to place an order for the replacement Model/Part Numbers (as referenced per #6. Per the FAQ's) please contact Vyaire Medical's Customer Service Department at 800.323.9088 (Option #1, followed by Option #1) Monday-Friday 8:00 am CST - 5:00 pm CST. *Please have your original PO number available when calling to expedite the credit process. CUSTOMERS THAT HAVE PURCHASED THROUGH A DISTRIBUTOR: ~ Please work with your distributor to receive credit of your in-stock product affected by this recall. For any additional questions and support concerning this voluntary recall, please contact Clinical Risk Coordinator, at 827-757-0109 or [email protected].
Quantity in commerce
16,670 units in total
Distribution
US Nationwide Distribution
Product codes and lots
0001100127 0001116871 0001116872 0001124936 0001129065 0001129066 0001129067 0001129069 0001129070 0001131426 0001131428 0001134350 0001135400 0001136027 0001136028 0001136029 0001145151 0001145152 0001145153 0001157125

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0346-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0343-2018Class ITerminatedVoluntary: Firm initiated
Z-0344-2018Class ITerminatedVoluntary: Firm initiated
Z-0345-2018Class ITerminatedVoluntary: Firm initiated
Z-0346-2018Class ITerminatedVoluntary: Firm initiated