TRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.5 MM, LENGTH 12 MM, REF TRY-PS-45 12-S, Qty 2 TRYPTIK(R)...
SPINEART SA began a recall of TRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.5 MM, LENGTH 12 MM, REF TRY-PS-45 12-S, Qty 2 TRYPTIK(R) 2C-Plate Anterior Cervical Plate System is... on . FDA classified it as Class II and gives the reason as "Product labeling for TRYPTIK 2-C-PLATE Self Tapping screw, REF TRY-PS 40, Lot 3-5934 contains the sketch/drawing of a self-drilling screw". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0339-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Spineart SA
- Product
- TRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.5 MM, LENGTH 12 MM, REF TRY-PS-45 12-S, Qty 2 TRYPTIK(R) 2C-Plate Anterior Cervical Plate System is intended to be used for temporary stabilization of the cervical spine (C2-C7) during the development of solid spinal fusion in patients with instability caused by the following degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumor, spondylolisthesis, spinal stenosis, deformity (i.e., scoliosis, kyphosis, lordosis), pseudarthrosis, and failed previous fusions.
- Product code
- KWQ Appliance, Fixation, Spinal Intervertebral Body
- Reason for recall
- Product labeling for TRYPTIK 2-C-PLATE Self Tapping screw, REF TRY-PS 40, Lot 3-5934 contains the sketch/drawing of a self-drilling screw.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Spineart sent an Urgent Medical Device letter by email dated November 27, 2017. The letter stated the following: "Out of an abundance of caution", we have decided to recall the concerned products. Please carefully follow instructions detailed below: 1. Immediately examine your inventory and quarantine product subject to recall. 2. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall by using this document. 3. Collect and quarantine all products. 4. Send back all products and the enclosed Response Form to Spineart Geneva E-mail: [email protected]." For further questions, please call (949) 313-6400..
- Quantity in commerce
- 9 units
- Distribution
- US Distribution to the state of : CA
- Product codes and lots
- batch: 3-5936
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0336-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0337-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0338-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0339-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0340-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0341-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0342-2018 | Class II | Terminated | Voluntary: Firm initiated |