HydroSil Gripper Intermittent Catheters- 10" lengths - 30 catheters per box, each catheter is packaged in an...
C.R. Bard, Inc. began a recall of HydroSil Gripper Intermittent Catheters- 10" lengths - 30 catheters per box, each catheter is packaged in an individual pouch. MGC3 Product Usage Intermittent... on . FDA classified it as Class II and gives the reason as "Several instances of a catheter caught in the sterile barrier seal and in some instances cut catheters were observed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1307-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- C.R. Bard Inc
- Product
- HydroSil Gripper Intermittent Catheters- 10" lengths - 30 catheters per box, each catheter is packaged in an individual pouch. MGC3 Product Usage Intermittent catheters are intended for urinary bladder drainage in patients requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures.
- Product code
- EZD Catheter, Straight
- Reason for recall
- Several instances of a catheter caught in the sterile barrier seal and in some instances cut catheters were observed.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A notification letter was sent to four hundred and fourteen (414) consignees via Federal Express on September 15, 2017. The letter explained the problem and requested return of the product. Distributors were directed to cease distribution of the recalled products and notify their customers of the recall.
- Quantity in commerce
- 44640.0 units
- Distribution
- US Nationwide in the states of AK, AL, AR, AZ, CA, FL, GA, HI, IA, ID, KS, KY, MD, ME, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, including Puerto Rico
- Product codes and lots
- Item 50610 Lots #'s: JUBT0404, JUBP0700, JUBR2498, JUBS1289
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1307-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1308-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1309-2018 | Class II | Terminated | Voluntary: Firm initiated |