FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The...

FUJIFILM Healthcare Americas Corporation / initiated 2017-06-15 / Terminated / root cause: design

Fujifilm Medical Systems U.S.A., Inc. began a recall of FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate... on . FDA classified it as Class II and gives the reason as "A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0232-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.
Product code
KPR System, X-Ray, Stationary
Reason for recall
A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.
Root cause tag
design
FDA root cause
Device Design
Action
FUJIFILM Medical Systems U.S. A. Inc. (FMSU) sent "Urgent: Medial Device Correction" recall notification letter dated June 21, 2017 to all Customers. Customers were notified of the recall, potential issue, potential hazard, description of the issue, the actions planned to correct the issue, the actions to be taken by the customer and contact information. A Customer Feedback Form was provided accompanying the letter. FMSU service personnel will contact all of the medical facilities where the applicable products have been installed to arrange for this correction and visit to take corrective measures. As FMSUs planned field correction at the affected sites, FMSU will install the additional protective parts preventing the X-ray tube assembly from coming into contact with the patient or operator should separation of the tube mounting flange from the tube holding shaft occur. For further questions, please call (203) 324-2000.
Quantity in commerce
3
Distribution
US Distribution.
Product codes and lots
UDI: (01)04540217052226(21)MP95A9F6B001; (01)04540217052226(21)MP95A9F6A001; (01)04540217052226(21)MP95A8F5A001.  Serial Numbers: FDR Visionary Suite System S/N MP95A9F6B001 MP95A9F6A001 MP95A8F5A001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0232-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0232-2018Class IITerminatedVoluntary: Firm initiated