Z-0515-2018Class II

ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810...

Siemens Healthcare Diagnostics Inc. / initiated 2017-06-07 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc began a recall of ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756 on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0515-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.
Product code
CHP Radioimmunoassay, Estradiol
Reason for recall
Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020).
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
An Urgent Medical Device Recall letter dated June 7, 2017 was distributed to all customers in the United States and an Urgent Field Safety Notice was distributed to all customers outside the United States who have received ADVIA Centaur Systems Calibrator 30 kit lot ending in 21(C3021). These letters instruct customers who operate the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems to discontinue use of C3021 . The letters inform customers that Calibrator 30 kit lots ending in 20 (C3020) and 23 (C3023) are not impacted and can be used. The letters inform customers who operate ADVIA Centaur CP systems to continue using their inventory of C3021, as they are not impacted by this issue. All affected domestic customers were sent a hard copy of the Urgent Medical Device Recall via Federal Express. Fax-back forms will be used to confirm that the customers received the Urgent Medical Device Recall letter.
Quantity in commerce
5992 units (US and OUS)
Distribution
US and OUS.
Product codes and lots
Lot Numbers: 00387A21, 16673A21, 36243A21;  00388A21, 15448A21, 51211A21;  20640A21; 51024A21.  Expiration Date: 2018-03-08

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0515-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0515-2018Class IITerminatedVoluntary: Firm initiated