Z-0234-2018Class II

Elekta ERGO++

Elekta Limited / initiated 2017-09-01 / Terminated / root cause: manufacturing process

Elekta Inc began a recall of Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments on . FDA classified it as Class II and gives the reason as "Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0234-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Elekta sent an Urgent Important Field Safety Notification letter dated September, 2017. The letter identified affected product, problem and actions to be taken. Customers were instructed that a copy of the Important Field Safety Notice is to be kept with the most current labeling and all personnel working with the product should be made aware of the content of the letter. The customers were asked to complete the verification tests and return the Important Field Safety Notice Acknowledgment form to their local Elekta representative as soon as possible but no later than 30 days from receipt. For questions contact your local Elekta Care Support Center.
Quantity in commerce
182
Distribution
Worldwide Distribution - US Nationwide in the states of: VA, CA, TX, MO, NY, WA, NM, MA, FL, CT, OK, SC, MS, NY, GA, NC and the countries of: AUSTRALIA, BANGLADESH, BRAZIL, BULGARIA, CANADA, COLOMBIA, CZECH REPUBLIC, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ITALY, JAPAN, JORDAN, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, PERU, PHILIPPINES, PORTUGAL, RUSSIA, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, TAIWAN, THAILAND, TURKEY, VENEZUELA, and VIETNAM
Product codes and lots
Version 1.7.3 and higher

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0234-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0234-2018Class IITerminatedVoluntary: Firm initiated