Z-0724-2018Class II

Integrity R1.2

Elekta Limited / initiated 2017-11-20 / Terminated / root cause: software

Elekta Limited began a recall of Integrity R1.2 on . FDA classified it as Class II and gives the reason as "Potential for positional errors following automatic table movement (ATM)". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0724-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Integrity R1.2
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Potential for positional errors following automatic table movement (ATM).
Root cause tag
software
FDA root cause
Software Design Change
Action
Customers were notified via letter on about [DATE]. Instructions include to post the notice in a place accessible to all users until the action is closed and to advise appropriate personnel working with the affected product of the recall action. Additionally Elekta plans to release Integrity R4.0.0, which should identify positional errors over 5 mm, in the first half of 2018.
Quantity in commerce
539 units total
Distribution
Distributed domestically to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed internationally to ALBANIA, ALGERIA, ANTIGUA AND BARBUDA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINIA, BOTSWANA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CUBA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, LAOS, LITHUANIA, MALAYSIA, MALTA, MEXICO, MOROCCO, MYANMAR, NAMIBIA, NEPAL, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA SUDAN, SURINAME, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UK, UKRAINE, VENEZUELA, VIETNAM, ZIMBABWE
Product codes and lots
Product code 90I--YE, no serial number

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0724-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0723-2018Class IITerminatedVoluntary: Firm initiated
Z-0724-2018Class IITerminatedVoluntary: Firm initiated