Z-0524-2018Class II

Syngo.plaza systems with SW VB20A, Model Number - 10863171, 10863172, 10863173

Siemens Medical Solutions USA, Inc. / initiated 2017-08-23 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo.plaza systems with SW VB20A, Model Number - 10863171, 10863172, 10863173 Product Usage: Syngo.plaza is a Picture Archiving and Communication System... on . FDA classified it as Class II and gives the reason as "When a prior study is being replaced in the workflow step, in certain scenarios (based on the Display Protocol configuration) the prior study will only be replaced in the active Workflow Step and not show in all other...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0524-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo.plaza systems with SW VB20A, Model Number - 10863171, 10863172, 10863173 Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
When a prior study is being replaced in the workflow step, in certain scenarios (based on the Display Protocol configuration) the prior study will only be replaced in the active Workflow Step and not show in all other workflow steps. The other workflow steps will continue to show the initially loaded study.
Root cause tag
software
FDA root cause
Software design
Action
Siemens Healthineers distributed Urgent Customer Advisory Notices to their customers via courier service. Customers were advised to always verify date and time of the displayed study during workflow step change to ensure that the desired images are being displayed. Otherwise, please replace the displayed initial study in the workflow step with the expected study. Customers with any questions, call Siemens Customer Care Centre at 888-826-9702.
Quantity in commerce
76
Distribution
US Nationwide Distribution
Product codes and lots
Model Number - 10863171, 10863172, 10863173

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0524-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0524-2018Class IITerminatedVoluntary: Firm initiated