Syngo.plaza systems with SW VB20A, Model Number - 10863171, 10863172, 10863173
Siemens Medical Solutions USA, Inc began a recall of Syngo.plaza systems with SW VB20A, Model Number - 10863171, 10863172, 10863173 Product Usage: Syngo.plaza is a Picture Archiving and Communication System... on . FDA classified it as Class II and gives the reason as "When a prior study is being replaced in the workflow step, in certain scenarios (based on the Display Protocol configuration) the prior study will only be replaced in the active Workflow Step and not show in all other...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0524-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Syngo.plaza systems with SW VB20A, Model Number - 10863171, 10863172, 10863173 Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- When a prior study is being replaced in the workflow step, in certain scenarios (based on the Display Protocol configuration) the prior study will only be replaced in the active Workflow Step and not show in all other workflow steps. The other workflow steps will continue to show the initially loaded study.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Healthineers distributed Urgent Customer Advisory Notices to their customers via courier service. Customers were advised to always verify date and time of the displayed study during workflow step change to ensure that the desired images are being displayed. Otherwise, please replace the displayed initial study in the workflow step with the expected study. Customers with any questions, call Siemens Customer Care Centre at 888-826-9702.
- Quantity in commerce
- 76
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Model Number - 10863171, 10863172, 10863173