Z-0486-2018Class II

Radial Artery Catheterization Set, 1) NA-04550-1A & 2) NA-04550-X1A, device permits access to the peripheral...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2017-07-10 / Terminated / root cause: manufacturing process

Arrow International Inc began a recall of Radial Artery Catheterization Set, 1) NA-04550-1A & 2) NA-04550-X1A, device permits access to the peripheral arterial circulation or to other small vessels on . FDA classified it as Class II and gives the reason as "Needle supplier indicated the presence of loose polypropylene foreign matter above release specification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0486-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Radial Artery Catheterization Set, 1) NA-04550-1A & 2) NA-04550-X1A, device permits access to the peripheral arterial circulation or to other small vessels.
Product code
FMI Needle, Hypodermic, Single Lumen
Reason for recall
Needle supplier indicated the presence of loose polypropylene foreign matter above release specification.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
On July 10, 2017, Arrow International distributed Urgent Medical Device Recall notice and acknowledgment forms to customers. Customers are advised to take the following actions: 1. If you have affected stock, immediately discontinue use and quarantine any products listed in the notice, so that the affected products can be returned to Arrow. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419- 8507, Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document your receipt of this letter. If you have any other questions, feel free to contact your local sales representative or Customer Service at 18662466990.
Distribution
...
Product codes and lots
Lot Numbers: 1) 23F16H0859 & 2) 23F16J0547

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0486-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0486-2018Class IITerminatedVoluntary: Firm initiated