10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400
Horiba Instruments, Inc dba Horiba Medical began a recall of 10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400 on . FDA classified it as Class II and gives the reason as "The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0493-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Product
- 10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm initiated their recall by letter on 11/21/2017. The letter stated the following: "A new production of reagent container compliant for each reference is under process and will be made available very soon. In the meantime and to allow a proper operating of the analyzer ABX PENTRA 400 / PENTRA C400, it Is necessary to push the reagent containers in the reagent racks until they touch the bottom of the rack. Please share this information with your laboratory staff, and retain this notification as part of your Quality System documentation. It is mandatory for you to complete and return the enclosed response form within 10 days so we may maintain our records."
- Quantity in commerce
- 10 bags, 100 units each
- Distribution
- Nationwide
- Product codes and lots
- DATE - MVT 25.08.2017
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0493-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0494-2018 | Class II | Terminated | Voluntary: Firm initiated |