Z-0605-2018Class II

Lyophilized Chemistry Calibrator Level 1

Siemens Healthcare Diagnostics Inc. / initiated 2017-08-24 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of Lyophilized Chemistry Calibrator Level 1 Product Usage: Beckman Coulter Chemistry Calibrators are intended for use when calibrating methods run on the Beckman... on . FDA classified it as Class II and gives the reason as "Unpaired lots of AU lyophilized chemistry calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. If lot numbers are...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0605-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Lyophilized Chemistry Calibrator Level 1 Product Usage: Beckman Coulter Chemistry Calibrators are intended for use when calibrating methods run on the Beckman Coulter AU series of chemistry analyzers. Beckman Coulter Chemistry Calibrators are lyophilized, human serum based products formulated for use as a reference material when calibrating Beckman Coulter AU clinical chemistry system assays. The calibrators are used to calibrate the following assays: Albumin (ALB), Bicarbonate (CO2), Direct Bilirubin (DBIL), Total Bilirubin (TBIL), Calcium Arsenazo (CALA), Calcium (CAL oCPC), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Phosphorus (PHOS), Iron, Lactate, Magnesium (MG), Total Protein (TP), Triglyceride (TRIG), Unbound Iron Binding Capacity (UIBC), Urea Nitrogen (BUN), and Uric Acid. Only lots 6101K51 and 6101K61 were subject to recall.
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
Unpaired lots of AU lyophilized chemistry calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results. QC failures could occur if incorrect pairs of calibrators are used.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
On August 28, 2017 a letter was sent to all the customers stating the following: Attention Beckman Coulter Customer, Beckman Coulter is initiating a field action for the product listed above. This letter contains important information that needs your immediate attention. ISSUE: Unpaired lots of AU Lyophilized Chemistry Calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. Paired calibrator lots are identified by the last two digits of the lot number. Calibrator Level 1 REF DR0070-1, Lot 6101K51 Calibrator Level 2 REF DR0070-2, Lot 6102K61 Incorrect pair Calibrator Level 1 REF DR0070-1, Lot 6101K61 Calibrator Level 2 REF DR0070-2, Lot 6102K61 Correct pair IMPACT: If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results. QC failures could occur if incorrect pairs of calibrators are used. Calibrator Level 1 REF DR0070-1, Lot 6101K51 Calibrator Level 2 REF DR0070-2, Lot 6102K61 Incorrect pair If the above incorrect pair is run with Level 1 lot 6101K51and Level 2 lot 6102K51 set points, erroneous results could be generated for the following assays: o Bicarbonate : maximum 11% bias low o Calcium Arsenazo : maximum 5% bias low o Calcium oCPC : maximum 5% bias low o Inorganic Phosphorous : 6.4% bias high o Magnesium: 3.2% bias low If the above incorrect pair is run with Level 1 lot 6101K61 and Level 2 lot 6102K61 set points, erroneous results could be generated for the following assays: o Bicarbonate : maximum 15.8% bias high o Direct Bilirubin OSR6x181: 8.7% bias low ACTION: Check the level 1 and level 2 calibrators to make sure you are using correctly paired lots for calibration. Ensure correct set points are being used for the lot. Set points are in the IFU for each lot and can be found in each kit, or on the Beckman Coulter websites
Quantity in commerce
unknown
Distribution
Worldwide Distribution in the states of AL, AK, AZ, AK, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IO, KS, KY, LA, MA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NY, NM, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VI, WA, WV, WI, and WY and the countries of Brazil, Canada, Costa Rica, El Salvador, Guam, Paraguay, and Peru
Product codes and lots
Lots # 6101K51 and 6101K61

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0605-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0605-2018Class IITerminatedVoluntary: Firm initiated