Z-0348-2018Class II

ECG Out Cable - Heartstart MRx M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A...

Philips North America LLC / initiated 2017-07-17 / Terminated / root cause: design

Philips Electronics North America Corporation began a recall of ECG Out Cable - Heartstart MRx M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables... on . FDA classified it as Class II and gives the reason as "Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0348-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
ECG Out Cable - Heartstart MRx M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641. Manufactured - September 4, 2002 October 8, 2014
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs): " On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave. " On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*. *Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obtaining the ECG signal from a bedside monitor. The XL+ Instructions for Use includes the following warning: When pacing in Demand Mode, the ECG cable from the patient must be directly connected to the HeartStart XL+. If the user follows this warning, this problem cannot occur on the XL+.
Root cause tag
design
FDA root cause
Device Design
Action
Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: Customers should remove affected sync cables and ECG out cables from service and destroy them, documenting such action on the attached reply form, which should be returned to Philips. Customers should retain a copy of the reply form. The customer should follow the Instructions for Use that describe the recommended approach for performing cardioversion with the XL+ and MRx monitor/defibrillators. The ECG trunk cable can be disconnected from the Philips bedside monitor and connected to the monitor/defibrillator before delivering therapy; this is the preferred source of an ECG waveform for synchronization.
Quantity in commerce
15811
Distribution
1067 US customers 163 Canada, 917 International
Product codes and lots
M1783A/M5526A 12-pin Sync cables  All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0348-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0348-2018Class IITerminatedVoluntary: Firm initiated