510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0269-2018Class II

ExploR Radial Head Replacement System. PN: 11-210061 LN: 243610. The stem is placed inside a polyurethane...

Zimmer GmbH / initiated 2017-06-16 / Terminated / root cause: packaging

Zimmer Biomet, Inc. began a recall of ExploR Radial Head Replacement System. PN: 11-210061 LN: 243610. The stem is placed inside a polyurethane bag. The bagged component will then be placed in an... on . FDA classified it as Class II and gives the reason as "Zimmer Biomet is conducting a medical device recall for the ExploR 6x24mm Implantable Stem with Screw due to a complaint alleging that the Modular Radial HD lock screw was not included in the package for the stem". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0269-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ExploR Radial Head Replacement System. PN: 11-210061 LN: 243610. The stem is placed inside a polyurethane bag. The bagged component will then be placed in an inner PETG tray. The screw will be placed inside a separate polybag, and placed in the PETG tray with foam pieces added to the top and bottom of the tray for protection. The inner tray and the contents within will then have a Tyvek lid heat sealed to the tray. The sealed inner tray will be fit into an outer PETG tray and a Tyvek lid heat sealed to the tray. The entire sealed unit is placed in a SBS carton that will be shrink wrapped for protection.
Product code
KWI Prosthesis, Elbow, Hemi-, Radial, Polymer
Reason for recall
Zimmer Biomet is conducting a medical device recall for the ExploR 6x24mm Implantable Stem with Screw due to a complaint alleging that the Modular Radial HD lock screw was not included in the package for the stem.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Zimmer Biomet sent US Fiedl action notices 6/16/17 via email and fed Ex. On 9/5/17, firm expanded their recall to include 1 lot and distributed Amended notices via email and FedEx
Distribution
US, Canada, OUS.
Product codes and lots
Part Lot UDI Number 11-210061 243610 and 202900; (01)00880304210448(17)260817(10)243610

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0269-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0269-2018Class IITerminatedVoluntary: Firm initiated