Z-0272-2018Class II
3040-VenaFlow Calf Garment (Aircast) Quick Connect was mislabeled as 3010-PL Aircast Pos-Lock Connector...
Stryker Sustainability Solutions / initiated 2017-12-01 / Terminated / root cause: packaging
Stryker Sustainability Solutions began a recall of 3040-VenaFlow Calf Garment (Aircast) Quick Connect was mislabeled as 3010-PL Aircast Pos-Lock Connector. Intended for use in preventing deep vein thrombosis... on . FDA classified it as Class II and gives the reason as "3040 compression sleeves being mislabeled as a 3010-PL compression sleeves". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0272-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Sustainability Solutions
- Product
- 3040-VenaFlow Calf Garment (Aircast) Quick Connect was mislabeled as 3010-PL Aircast Pos-Lock Connector. Intended for use in preventing deep vein thrombosis (DVT), diminishing post-operative pain and swelling, enhancing blood circulation and reducing wound healing time on the calf.
- Product code
- JOW Sleeve, Limb, Compressible
- Reason for recall
- 3040 compression sleeves being mislabeled as a 3010-PL compression sleeves.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- An Urgent Medical Device Recall letter, dated December 5, 2017, was distributed to customers. The letter described the affected product and stated the reason for the recall, as well as the risk to health. Customers were instructed to discontinue use of the affected device ,and complete and return the recall effectiveness check form. All affected product should be returned per the instructions provided and a credit will be issued.
- Quantity in commerce
- 280 devices
- Distribution
- Distributed to Texas.
- Product codes and lots
- Lot 6285035. UDI: 00885825016661.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0272-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY