Z-0327-2018Class II

EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic...

Encore Medical, LP / initiated 2017-12-12 / Terminated / root cause: design

Encore Medical, Lp began a recall of EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic surgery on . FDA classified it as Class II and gives the reason as "The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0327-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic surgery.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.
Root cause tag
design
FDA root cause
Device Design
Action
The field safety notice was disseminated by email on 12/12/2017. The notice stated the following: "DJO Sales Agents Action: Please review the bulletin that is provided with this notice. This bulletin contains new instructions on the use of the Locking Punch Guide and Punch. Please ensure your surgeons are familiar with this information. DJO Surgical then requires that you complete the attached acknowledgment and email it to [email protected]. Please have this form returned by December 15, 2017 Once redesigned guides are available, they will be exchanged through a separate field action."
Quantity in commerce
2 units
Distribution
One medical device distributor in California.
Product codes and lots
Lot 252564

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0327-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0326-2018Class IITerminatedVoluntary: Firm initiated
Z-0327-2018Class IITerminatedVoluntary: Firm initiated
Z-0328-2018Class IITerminatedVoluntary: Firm initiated
Z-0329-2018Class IITerminatedVoluntary: Firm initiated