Philips SureSigns VSi / SureSigns VS2+ vital signs monitor used in healthcare environments and transport...
Philips Electronics North America Corporation began a recall of Philips SureSigns VSi / SureSigns VS2+ vital signs monitor used in healthcare environments and transport within a facility on . FDA classified it as Class III and gives the reason as "Some SureSigns VSi, VS2+ products have the Date of Manufacture on its serial number label as %DoM , instead of the specific date in the format of YYYY-MM". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0268-2018
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips SureSigns VSi / SureSigns VS2+ vital signs monitor used in healthcare environments and transport within a facility.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Some SureSigns VSi, VS2+ products have the Date of Manufacture on its serial number label as %DoM , instead of the specific date in the format of YYYY-MM.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Philips sent each known affected customer a Field Safety Notice, dated September, 2017. The Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips will provide a corrected label free of charge
- Quantity in commerce
- 1079
- Distribution
- Worldwide distribution. Australia, Austria, Bahrain, Belgium, Brazil, Denmark, France, French Polynesia, Germany, Hong Kong, India, Ireland, Israel, Italy, Kenya, Lebanon, Lesotho, Malaysia, Maldives, Morocco, Netherlands, Oman, Pakistan, Palestine, Poland, Saudi Arabia, South Africa, Spain, Sweden, Thailand, Turkey, United Kingdom, UAE, Vietnam, and Zimbabwe.
- Product codes and lots
- SureSigns VSi : 863275, 863276, 863277 SureSigns VS2+ : 863278, 863279
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0268-2018 | Class III | Terminated | Voluntary: Firm initiated |