IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Measurement...
Philips Electronics North America Corporation began a recall of IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Measurement Server X2, IntelliVue Patient Monitor Models MP2... on . FDA classified it as Class II and gives the reason as "Distorted ST-Segment when using 12-Lead ECG Monitoring with Philips IntelliVue Patient Monitors". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0292-2018
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Measurement Server X2, IntelliVue Patient Monitor Models MP2 and MP5 Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Distorted ST-Segment when using 12-Lead ECG Monitoring with Philips IntelliVue Patient Monitors
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips sent an Urgent Medical Device Correction letter to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Philips will provide software update and is provding interim instructions for the operator. For questions contact your local Philips representative.
- Quantity in commerce
- 7071
- Distribution
- Worldwide Distribution and US Nationwide
- Product codes and lots
- Model #: M300IA/AL, M3002A, M8102A, M8105A with ECO FW revision E.0 1.22 and 12-lead option C#12 Lots #s included on Consignee list Catalog #: 862442 (M3001A), 866462 (M3001AL), 865039 (M3002A), 865040 (M8102A), 865024 (M8105A)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0292-2018 | Class II | Ongoing | Voluntary: Firm initiated |