510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0499-2018Class II

Roche / Hitachi MODULAR Analyzer Systems-Immunoassay Analyzer-Model/Catalog/Part Number...

Roche Diagnostics GmbH / initiated 2017-07-05 / Terminated / root cause: software

Roche Diagnostics Corporation began a recall of Roche / Hitachi MODULAR Analyzer Systems-Immunoassay Analyzer-Model/Catalog/Part Number: 11568248001/11568248692 - Elecsys 2010 analyzer (disk system) and... on . FDA classified it as Class II and gives the reason as "A software malfunction can occur on the cobas e 411 and Elecsys 2010 analyzers in the Sample& Control data file which may lead to a potential data mismatch". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0499-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche / Hitachi MODULAR Analyzer Systems-Immunoassay Analyzer-Model/Catalog/Part Number: 11568248001/11568248692 - Elecsys 2010 analyzer (disk system) and 11804014001/11804014692 - Elecsys 2010 analyzer (rack system) Medical Device Listing number: D053244 - Elecsys 2010 Elecsys analyzer is a fully automated, random access, computer controlled analytical systems for quantitative and qualitative determinations of analytes in body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
A software malfunction can occur on the cobas e 411 and Elecsys 2010 analyzers in the Sample& Control data file which may lead to a potential data mismatch.
Root cause tag
software
FDA root cause
Software design
Action
All consignees were notified by Urgent Medical Device Correction sent via UPS Ground (receipt required) on July 5, 2017. The customer is advised to perform the following actions: cobas e 411: Until the software update is available, use the workaround outlined in the Urgent Medical Device Correction (UMDC). Elecsys 2010: Use the workaround outlined in this Urgent Medical Device Correction (UMDC). Non-responding accounts will be defined as users for which confirmation of receipt of the Urgent Medical Device Correction (UMDC) cannot be obtained. These accounts will be monitored on an ongoing basis and follow-up attempts will be made accordingly. Workaround for customers: he customer is advised to perform the following actions: cobas e 411: Until the software update is available, use the workaround outlined in the Urgent Medical Device Correction (UMDC). Elecsys 2010: Use the workaround outlined in this Urgent Medical Device Correction (UMDC). For further questions, please call (317) 521-7259
Quantity in commerce
255 instruments
Distribution
Nationwide Distribution
Product codes and lots
none

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0499-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0498-2018Class IITerminatedVoluntary: Firm initiated
Z-0499-2018Class IITerminatedVoluntary: Firm initiated