510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0492-2018Class II

TosohSorter 9tray, Product Code 022931and TosohSorter 19tray, Product Code 022932

Tosoh Bioscience Inc / initiated 2017-06-11 / Terminated / root cause: manufacturing process

Tosoh Bioscience Inc began a recall of TosohSorter 9tray, Product Code 022931and TosohSorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports... on . FDA classified it as Class II and gives the reason as "This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0492-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Tosoh Bioscience Inc
Product
TosohSorter 9tray, Product Code 022931and TosohSorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.
Product code
KHO Fluorometer, For Clinical Use
Reason for recall
This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Customers were sent a recall notification by email on July 11, 2017 informing them that their local Field Service Engineer would contact them to make arrangements to proactively replace the affected parts. Follow up notifications will be by Federal Express with verification of delivery.
Quantity in commerce
565 total devices U.S.
Distribution
Worldwide Distribution - U.S. Nationwide - Latin America
Product codes and lots
AIA-900 19tray Sorter with S/N 10000109 to S/N 80136909. AIA-900 9tray Sorter with S/N 10000109 to S/N 10707402.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0492-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0491-2018Class IITerminatedVoluntary: Firm initiated
Z-0492-2018Class IITerminatedVoluntary: Firm initiated