Unifine Pentips 5mm x 31G pen needles, packaged in 30 ct pen needles
Owen Mumford USA, Inc. began a recall of Unifine Pentips 5mm x 31G pen needles, packaged in 30 ct pen needles. Product Usage: Sterile, non-toxic, non-pyrogenic, single use pen needles for the... on . FDA classified it as Class II and gives the reason as "Device may potentially be compromised by water in end user packaging. Exposure to liquid may compromise sterility. This may appear as discoloring or yellowing of the device's protective seal". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0382-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Owen Mumford USA, Inc.
- Product
- Unifine Pentips 5mm x 31G pen needles, packaged in 30 ct pen needles. Product Usage: Sterile, non-toxic, non-pyrogenic, single use pen needles for the subcutaneous injection of drugs from pen injectors.
- Product code
- FMI Needle, Hypodermic, Single Lumen
- Reason for recall
- Device may potentially be compromised by water in end user packaging. Exposure to liquid may compromise sterility. This may appear as discoloring or yellowing of the device's protective seal.
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- Owen Mumford sent an Urgent Medical Device Recall Notification letter dated November 6, 2017 via email. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to check inventory for affected product, contact the Customer Service Department to arrange return of affected product, complete and return the response form and to either provide a customer list for Owen and Mumford to notify their customers or confirmation that they will notify their customers if further distributed.
- Quantity in commerce
- 708 cartons (21,240 pen needles)
- Distribution
- US Nationwide Distribution in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, MA, MI, MN, MO, MS, NC, NH NJ, NY, OH, OK, PA, TN, NX, UT, VA, VT, WA, WV
- Product codes and lots
- UPC #384701150012: Batch No. AN 1150, Lot No.1705711.T098; Batch No. 1150REC, Lot No. 1705722.T099; Batch No. AN 1150HEB, Lot No. 1705722.T100.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0381-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0382-2018 | Class II | Terminated | Voluntary: Firm initiated |